LOG-IN
Displaying reports 901-920 of 3201.Go to page Start 42 43 44 45 46 47 48 49 50 End
KAGRA MIR (Polarization)
Print this report.
MarcEisenmann - 20:43, Tuesday 14 September 2021 (2681)Get code to link to this report
birefringence setup almost recovered

Marc, Matteo

 

Today we first removed the PBS (it was taking too much place and prevented the installation of QWP and HWP).

We tweaked the last lens on the injection (L2) and recovered proper signals using the surface reference sample.

We placed the pinhole and tune its position at the waist (z=41mm) and plus/minus one Rayleigh range (ie Z = 36 and 46 mm). We checked that at these position all the pump power was properly transmitted (ie P_t = P_in = 28.8 mW)

We installed back the PBS and started to try to align it.

It was not possible to recover more than 22 mW. We checked that it was not due to misalignment by moving the pinhole that also gave 22 mW of transmitted power at maximum.

We simulated the effect of the PBS on waist size and position using Jammt : It moves the waist by ~ 10 cm. But this can be recovered by moving L2 by ~1 cm towards the imaging unit.

We did this change and could recover more than 27 mW at the extremum positions of the pinhole.

We then placed the razor blade (that cuts the beam in vertical) so that it is at the center of the pinhole holder.

We wanted to use the Translation_Stage_v3.vi to automatize the measurement but no values except the position changed...

By using the ..._v2.vi we got the beam waist = 33.39 um at z = 40.76 mm (see figure 1).

We reinstalled the surface reference sample and moved the translation stage to 40.8+1.5 = 42.3 mm (to compensate for the 3 mm thickness of this sample) and by tuning the horizontal and vertical screws of L2 we recovered the expected AC value (ie >0.4V).

We reinstalled the QWP and HWP, beam dump in reflection of the PBS and a small one instead of the steering mirror that catches the pump beam reflection from the sample (not enough space with the current holder) as in figure 2.

We did a Z scan of this sample and found R = 18.57 /W.

Either the alignment was the best since few months, either there are some polarizations effects that prevented to reach this value before I guess.

Finally, we started to reinstall the imaging unit part as in figure 3.

The steering mirror that was removed is now placed in front of the high-power power meter to allow for quick switch between absorption and birefringence measurement.

A 2" lens and the PBS inside the black tubing were also installed.

We still have to install ND filters and the 2 photodiodes on the readout path as well as proper walls.

Also, the motorized HWP is not yet connected to PC.

Images attached to this report
2681_20210914133521_pumpbeam.jpg 2681_20210914134335_img20210914195540.jpg 2681_20210914134340_img20210914195548.jpg
R&D (FilterCavity)
Print this report.
MichaelPage - 15:40, Friday 10 September 2021 (2679)Get code to link to this report
Retrying new Mephisto RIN measurement, comparison with other results

Yuhang and Michael

We measured again the RIN of the new Mephisto laser. The laser was sent to a tilted ND filter (prevents back reflection) -> 75mm focal length lens -> PD. The measurements were taken with the spectrum analyser taken out of the clean room.

The relative intensity noise and PD dark noise are shown in figures 1 and 2 (low freq span, high freq span). The units differ between the curves, though the RIN isn't recalibrated by much - the intensity noise was measured from the PD and then divided by the DC value of 0.89 V. For comparison, the KAGRA HP laser RIN shown in Aso-san's LVK presentation is given in figure 3. The new Mephisto laser at TAMA seems to have better RIN than the "Current" KAGRA laser, but not quite as good as the "New" laser.

Images attached to this report
2679_20210910083355_rin0909.jpg 2679_20210910083359_rinhf0909.jpg 2679_20210910083501_kagrarin.png
KAGRA MIR (General)
Print this report.
MarcEisenmann - 15:51, Tuesday 07 September 2021 (2678)Get code to link to this report
start of modification for birefringence measurement

Today I checked the 2 PBS available in PCI.

There is one with label (1064 nm PBS) so I decided to use this one.

I moved it to the mount with the good height but I'll need to print a new label for it.

I measured its transmission to be 166.0 uW with 166.4 uW of incident power.

Actually, it seems that the beam is too large for the small power-meter head as the power is supposed to be 30 mW there...

Anyway I started to tweak the alignment by checking the AC value of the surface reference sample.

Up to now I did not manage to recover proper alignment and it seems that the shape of the AC signal during Z scan is a bit strange (kind of similar to bulk sample...)

Hopefully I'll have a bit more time tomorrow to finish this alignment.

Note that to save time the pump laser is kept on with low power (30 mW)

KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 11:34, Monday 06 September 2021 (2677)Get code to link to this report
Comment to SHINKOSHA eval plate #7 absorption measurement started (Click here to view original report: 2674)

I attached the measurement results which gives absorption = 267+/-59 ppm/cm with P_in = 3.745 W and P_t = 3.198W.

Images attached to this comment
2677_20210906043401_xy.png 2677_20210906043405_xz.png 2677_20210906043408_yz.png
R&D (FilterCavity)
Print this report.
MichaelPage - 17:07, Friday 03 September 2021 (2676)Get code to link to this report
OPO replacement - incoupler characterisation

Yuhang and Michael

We did a quick measurement of the new incoupling mirror using the OPO replacement setup in ATC. The incoupling mirror was placed with the convex side facing forward in the mirror mount. The beam was reflected at a low angle (approximately 10 degrees), which is within the specified incident angle tolerance. Reflection and transmission were measured on both faces of the incoupler.

Using thorlabs power meter, we obtained:

Forward mount

Incident =   6.58 mW
Reflected =  6.07 mW (92.2%)
Transmitted = 0.537 mW (8.1%)

Reverse Mount

Incident = 6.64 mW
Reflected = 6.04 mW (91.0%)
Transmitted = 0.535 mW (8.1%)

The power meter is imprecise when being intermittently moved around between measurements - certain beam spot positions around the edge of the power meter seem to record more power (also noted during a previous measurement of OPO nonlinear gain in TAMA)

Very rough measurement obtained by focusing ceiling lights onto the table indicates the focal length to be approximately 15cm.

KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 18:33, Wednesday 01 September 2021 (2674)Get code to link to this report
SHINKOSHA eval plate #7 absorption measurement started

Today I installed the SHINKOSHA evaluation plate #7 on the translation stage so that the top side of the ingot is facing the laser sources with marking on the right side of the sample looking from the laser source side.

I moved the IU to 62.6 mm to compensate for the thickness.

I checked the X and Y centering using drop of the DC values to ~ 8mV which gave : X_center = 397.466 mm and Y_center = 120.629 mm.

I increased the pump power to ~3W and checked the Z centering at that position from the phase = 0 deg and got Z_center = 53.2 mm.

I also checked the top/ bottom and left/right Z centering over a 120 mm diameter area and found a tilt of this sample in both these directions to be ~ 0.8 mm.

Matteo confirmed that it was small enough to start absorption measurement.

The XY absorption measurement started over a circular area with diameter 120 mm and step size 250 um with P_t = 3.198 W and P_in = 3.745 W

Comments related to this report
MarcEisenmann - 11:34, Monday 06 September 2021 (2677)

I attached the measurement results which gives absorption = 267+/-59 ppm/cm with P_in = 3.745 W and P_t = 3.198W.

R&D (FilterCavity)
Print this report.
YuhangZhao - 16:36, Wednesday 01 September 2021 (2673)Get code to link to this report
Zero baseline drift monitoring

Mainly two mechanisms cause zero baseline drift (ZBD): the birefringence of electro-optic crystal and the etalon effect formed by two parallel end facets of the EOM crystal [Z. Li, et al., Optics Letters, 41, 14, 2016]. The reduction of birefringence effect can be done by controlling modulation voltage or the crystal's temperature [K. Kokeyama, et al., J. Opt. Soc. Am. A 31, 81 (2014)].

To monitor ZBD, we can lock cavity on anti-resonance. In this situation, the PDH signal is very much insensitive to phase change. Therefore, the signal magnitude change comes mainly from sideband amplitude change, usually addressed as residual amplitude modulation (RAM). This amount of RAM exists all the time, including the situation when cavity is on resonance. The control loop makes error signal to be zero with such RAM present, which makes the cavity not exactly on resonance. Or we can say RAM introducing detuning for cavity. This is the effect we are interested in. Notably, this effect may appear also for the CCFC control loop, which should be investigated.

We used the method described in the last paragraph to monitor ZBD. Since this effect exists when cavity is on resonance, we use the slope of PDH signal at resonance to calibrate ZBD. In this way, we have the detuning influence cause by ZBD for detuning change. The attached figure (is not yet calibrated), after calibration, shows a ~4Hz detuning change around zero.

Images attached to this report
2673_20210901093635_zbd.png
KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 10:55, Wednesday 01 September 2021 (2672)Get code to link to this report
Comment to Tama sample T1 absorption measurement started (Click here to view original report: 2670)

Measurement finished and reported in the 3 attached figures.

I finally choose to use this colormap for all plots as it is (I think) nice looking, color-blind and printing friendly.

Mean absorption seems to be 18+/-5 ppm/cm.

The sample was oriented to have the kanji for 'up' at the top.

Images attached to this comment
2672_20210901035433_xy.png 2672_20210901035439_xz.png 2672_20210901035645_yz.png
KAGRA MIR (General)
Print this report.
YuhangZhao - 17:25, Tuesday 31 August 2021 (2671)Get code to link to this report
Two HEPA capsules installed for PCI cleanroom air gun

Two HEPA capsules are installed for the air gun of PCI cleanroom. The maximum air pressure was specified to be 3.4 bar (figure 1). So I adjusted the pressure of the compressed air as attached figure2.

History: Marc found that there is air leakage between the connection part of capsule and tube. The connection was done by only fixing capsule and tube with steel belt. Aso-san put some white sealing tapes in this connection part. The use of sealing tape solves the problem of air leakage.

Images attached to this report
2671_20210831102449_wechatimg756.jpeg 2671_20210831102457_wechatimg755.jpeg
KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 10:27, Tuesday 31 August 2021 (2670)Get code to link to this report
Tama sample T1 absorption measurement started

Yesterday I started the measurement of the tama size sample T1.

Thanks to the help of Simon and Matteo I could identify the good sample.

Indeed there are now 2 tama size sample in PCI clean room but one seems to be different (maybe FC mirror?)

The one I installed is the one on the left of entry 301.

However, it seems that this sample creates at least 2 beams in reflection that hit directly the last prism on the probe path before the sample.

In turn, multiple reflections are sent a bit everywhere (but not towards the photo detector).

Using 50 mW of incident pump power I could estimate the power lost in these reflections to be around 10% of the incident power.

Because the sizes of these reflections are quite large, we started the measurement anyway.

These measurements will finish this evening and I will turn off the pump laser at that time.

Comments related to this report
MarcEisenmann - 10:55, Wednesday 01 September 2021 (2672)

Measurement finished and reported in the 3 attached figures.

I finally choose to use this colormap for all plots as it is (I think) nice looking, color-blind and printing friendly.

Mean absorption seems to be 18+/-5 ppm/cm.

The sample was oriented to have the kanji for 'up' at the top.

KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 14:47, Monday 30 August 2021 (2669)Get code to link to this report
AZTEC tama size sample absorption

The 3 measurements (XY, XZ and YZ) have been finished and are reported on the attached figures.

Note that for the XZ and YZ measurements the mean and standard deviation are computed only inside the sample while the contour plot and histogram show the entire measurement.

The results is compatible with previous measurements and gives absorption ~ 60 +/- 5 ppm/cm without peculiar feature on XY plane.

I like a lot the plots in gray scale but I also did the same with colored scale.

I'll upload soon a notebook to do these plots.

The mirror is oriented to have a black dot on the top.

EDIT : changed number of bins in histogram

Images attached to this report
2669_20210830122012_xy.png 2669_20210830122015_xz.png 2669_20210830122020_yz.png 2669_20210830122029_xycolored.png 2669_20210830122035_xzcolored.png 2669_20210830122040_yzcolored.png
R&D (FilterCavity)
Print this report.
MichaelPage - 19:35, Thursday 26 August 2021 (2665)Get code to link to this report
New Mephisto laser at TAMA - initial relative intensity noise test

Yuhang, Michael

We measured the relative intensity noise of the new Mephisto 2W NdYAG laser to be used at TAMA. The laser was sent into a 75mm lens, then a 50:50 beamsplitter going to a photodetector and power meter sensor. The distance between the laser and detectors is fairly short due to the initial divergence of the laser beam. With the photodetector we took the following spectra, while the power meter was used to measure long term power stability. Some ND filters were placed between the laser and beam splitter to counteract saturation of the PD and prevent back reflection into the laser. 

First PD: Thorlabs PDA36A-EC 

Second PD: Qubic PD-AC100-Si 1-100 MHz

Figure 1 shows the low frequency laser RIN at approximately the minimum current required for the laser to start emitting (about 0.8 A), compared to the PD dark noise. The measurement was passed through the Stanford pre-amp at DC coupling and 100 gain.

Figure 2 shows the RIN at different laser current (power). The DC level on the detector changes on the highest two measurements due to the PD saturating. The gain on the pre-amp was reduced by a factor of half for the highest two current measurements. Due to the PD reading being close to the noise floor, this may be responsible for the jump between the levels of the measurements between different pre-amp gain values.

Figure 3 shows the effect of activating the noise eater. In our first measurement its effect only became noticeable at laser current > 1.55 A.

Figure 4 shows the predicted frequency dependence of RIN from the Coherent Mephisto manual. Ideally the noise eater suppresses the large peak and reduces the 10k-100kHz noise floor by ~15 dB. We don't see much if any of the latter effect though. It seems like the low frequency signal is too close to the detector noise level. The levels are a bit different from the measured results - I didn't apply normalization to the measurements from the spectrum analyzer readings, but the relative values for noise eater on/off are what is important.

To follow up, we tried changing the preamp gain settings and changing the spectrum analyser coupling from DC to AC to see if we could get better SNR. The Stanford pre-amp was close to saturating though.

Figure 5 shows the result of increasing the PD gain and optimising the pre-amp settings. We can get a bit more clearance from the PD dark noise versus figure 1, but still less than desireable.

We then tried to do another measurement using the second PD, but it seemed like it had more dark noise than the Thorlabs one just from inspection on the oscilloscope. The Qubic PD is AC only, but we couldn't see anything on the oscilloscope when applying 1-10 kHz modulation, just ~ 5 mV rms noise.

Images attached to this report
2665_20210826121031_darknoise0826.jpg 2665_20210826123832_lasercurrent0826.jpg 2665_20210826123837_noiseeater0826.jpg 2665_20210826123844_mephistoringiven.png 2665_20210826123852_pdgain.jpg
KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 16:50, Thursday 26 August 2021 (2668)Get code to link to this report
Kasi absorption measurement started

Today I checked the bulk reference and got : R_bulk = 0.7527 cm/W (note that in this case I moved the IU by 0.32 mm)

With laser diode current of 6A I set up the HWP so to have P_in = 7.072W.

Using the DC values drop, the Kasi sample center positions are : X_center = 329.62 mm and Y_center = 122.81 mm.

The Zaber limits have been changed accordingly with some margin to check the incident power from remote.

I did a long Z scan and used the phase transition to get the center Z position at Z_center = 64.1 mm.

I realigned the DC at this position and started a circular map in the XY plane with parameters :

X = X_center, Y=Y_center, Z =Z_center, 20 mm radius, 0.1 mm step size, 500 ms waiting time, median/average filter order 5 and sensitivity = 5mV (ie 5 nA).

Preliminary absorption is : 75 ppm/cm (coherent with previous measurement)

with :

AC =4.6e-4 V
DC = 4.144 V
P_in = 7.074 W
P_t = 6.096 W
T_sample = P_t/P_in;
mat_correction=3.34;

All measurements should be finished tomorrow.

KAGRA MIR (General)
Print this report.
MarcEisenmann - 17:54, Wednesday 25 August 2021 (2667)Get code to link to this report
Re alignment of the setup

Katsuki, Marc

We performed several scans of the the z positions of the translation stage or the IU with the surface reference sample.

We plan to check if the shape of these signals could be used to perform more efficiently the alignment.

We tried to place back the PBS for the birefringence measurement but it was quite lossy (~20%).

Actually the one we used does not have name but the holder height seems to match. So we installed it as in entry 1496.

I'm planning to buy a new PBS as I could not find other.

The R_surf is now 16.79 /W at z_translationStage = 40.5 mm and z_IU = 66 mm.

Tomorrow we'll start the absorption measurement of the korean sample.

R&D (FilterCavity)
Print this report.
YuhangZhao - 11:15, Wednesday 25 August 2021 (2666)Get code to link to this report
New Mephisto power stability

Michael and Yuhang

We received a new Mephisto laser this Monday. Michael and I tested lasing current threshold, noise eater engagement current threshold. We measured RIN with different frequency band, current, noise eater on/off, and power stability. In this entry, we report new Mephisto power stability.

We used a power meter to monitor the power stability of new Mephisto. The total monitoring time is 17.5 hours. The power change is attached in the figure. A peak to peak 1.4% power stability was found with irregular power change found at the end of measurement. This shows that laser power doesn't become really stable after long time operation.  

Images attached to this report
2666_20210825041236_13.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 18:05, Tuesday 17 August 2021 (2663)Get code to link to this report
IR detuning stability with GR lock investigation with AA loop open (but AA error signals are put around zero by hand)

The IR detuning map in elog2615 was concerned that it could be related with alignment control. Therefore, I try to clarify this concern here and do some related test.

How is the AA loop?

     As far as I found, the AA loop bring GR and IR transmission to top of TEM00. If I introduce pertubation to AA loop, GR power goes down, which means AA loop works well.

Why IR detuning map should not be related to AA?

    1. The IR detuning map in elog2615 should not be related to AA loop because the AA loop works to have a totally linear response as elog2650. But IR detuning map has a flat region.

    2. I took IR detuning spectrum with AA loop opened. Meanwhile, the AA error signals are put around zero by hand. The beam hitting position was chosen as elog2573 to old and new spot. Then I got two spectrums are the attached figure. The spectrum shows result as found in elog2573. Since AA loop is open, there should be not effect from AA loop.

Images attached to this report
2663_20210817110231_detuningnoaa.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 17:23, Tuesday 17 August 2021 (2660)Get code to link to this report
Filter cavity alignment recovery

It is found, as attached figure 2, that PR and BS mirror had a sudden position change on last Thursday (20210812). This makes the filter cavity alignment totally lost. Since the movement of PR mirror is so large that the PR mirror picomotor needed to be used to recover alignment.

The mirror movement has coincidence with an earthquake (figure 1). But it is the first time I notice that mainly PR pitch mirror is moved. BS is moved a bit as well. But no obvious mirror movement is found for input and end.

Images attached to this report
2660_20210817102248_earthquake.png 2660_20210817102349_prsuddenchange.png
General (General)
Print this report.
YuhangZhao - 17:13, Tuesday 17 August 2021 (2659)Get code to link to this report
The bicycles in south and west arms are exchanged

The tire of the bicycle in the south arm was broken. However, the one in the west arm was OK. So I exchanged the place of them. In the future, we can use a good bicycle (blue color as attached picture) to go to the end room of south arm.

Images attached to this report
2659_20210817101356_2449420449222702925520210817170816.jpg
R&D (FilterCavity)
Print this report.
NaokiAritomi - 18:39, Wednesday 11 August 2021 (2658)Get code to link to this report
CCFC error signal for 3 hours

I measured CCFC error signal for 3 hours with 30 minutes interval (Fig. 1). Fig. 2 shows the CCFC error signals around 0 crossing point. The 0 crossing point of the fitting result (dashed curves) changed by (1.04-0.97)*54 = 3.8 Hz in 3 hours.

I also measured the nonlinear gain and shot noise before/after the whole measurement.

The nonlinear gain changed from 4.7 to 4.3 in 3 hours. This nonlinear gain change causes the CCFC amplitude change. The normalized CCFC amplitude can be written as follows.

Normalized CCFC amplitude = x/(1-x^2)^2 = (1-1/sqrt(g))/(2/sqrt(g)-1/g)^2

Figure 3 shows the normalized CCFC amplitude as a function of nonlinear gain. When the nonlinear gain changes from 4.7 to 4.3, the normalized CCFC amplitude changes from 1.07 to 0.97 by a factor of 0.9. In fact, the CCFC amplitude changed from 132mVpp to 118mVpp by a factor of 0.9 in 3 hours.

Images attached to this report
2658_20210811113850_20210810ccfc.png 2658_20210811113858_20210810ccfcenlarge.png 2658_20210811113908_ccfcamplitude.png
R&D (FilterCavity)
Print this report.
NaokiAritomi - 14:50, Wednesday 11 August 2021 (2657)Get code to link to this report
CCFC error signal with different length LEMO cables for CCFC LO

I measured CCFC error signal with different length LEMO cables for CCFC LO (attached figure). The CCFC amplitude was 132mVpp. The red and green curves in the figure represent the CCFC error signal with red and brown+green LEMO cables for CCFC LO, respectively. As you can see, the CCFC error signal with brown+green LEMO cable is close to I phase. So I will use the brown+green cable for CCFC LO.

Images attached to this report
2657_20210811075010_ccfclemo20210810.png