LOG-IN
Displaying reports 1801-1820 of 3201.Go to page Start 87 88 89 90 91 92 93 94 95 End
R&D (FilterCavity)
Print this report.
YuefanGuo - 00:46, Wednesday 18 September 2019 (1645)Get code to link to this report
Re-measure the reflection green beam after 500mm lens

Yuefan and Yuhang

As planned last week, today we measured the reflection beam again after the 500mm lens with cavity locked. The fitting shows in the attached figure. 

Same as last time, the position of the waist in two axis are more or less the same, but they both moved around 8cm from 59cm from the lens to 51cm to the lens this time.

The waist size is very different. In last measurement the waist was 64um and 92um, this time is 60um and 119um.  This gives the reflection beam waist size and position.

1. 711um. 0.863m before the lens, which should be between the periscope and first mirror.

2. 1274um, 5 m before the lens.

The difference in these two measurements, we guess is caused by the alignment condition. Because when we installed the mirrors and lens last week, they were well centered, but today when we aligned the cavity, the beam looks obviously lower than the center of the lens, which will induce some more astigmatism. But since the lens has a focal length of 500mm, it didn't affect so much, but in the design the second lens has a focal length of -50mm, if everytime we realign the cavity, the beam is off center on that lens, it could be a problem.

The solution we were thinking about is to put one or two iris on the reflection beam path when we centered all the optics, then they will act as reference for future alignment.

Anyway, apparently the beam has astigmatism, which may bring some problem to the automatic alignment system. So tomorrow if it is possible, we will move the last lens on the green input path because it is really bad algined now.

Images attached to this report
1645_20190917165331_greenrefafter500mmlens.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 23:48, Tuesday 17 September 2019 (1644)Get code to link to this report
Filter cavity open loop transfer function measurement and GR filter cavity transmission

Yuefan and Yuhang

Recently, we always have a problem of filter cavity green transmission power drop. It dropped from 5000 to 1500 recently. We figured out this problem today although we are always checking with the same method. The reason for GRtra power drop is too high gain.

After reducing the gain of filter cavity control loop while looking at filter cavity green transmission, we found the indicator of GRtra goes to 5000.

The incident power to filter cavity is 12.5mW, setting of the filter cavity loop is attached in figure 1 and 2(attenuation:1.68, gain:2.28). The locking bandwidth is 21kHz.

Images attached to this report
1644_20190917164846_wechatimg524.jpeg 1644_20190917164852_wechatimg523.jpeg 1644_20190918170438_figure1.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 23:30, Tuesday 17 September 2019 (1643)Get code to link to this report
Galvo loop closed with beam jittering and astigmatism

Yuhang and Yuefan

Since now we have the reflected beam coming from the filter cavity, it has astigmatism of ratio almost 2 between two axes. And also, it was jittering while testing.

It is reasonable that the range is larger at yaw direction while smaller at pitch direction due to astigmatism.

Anyway, the galvo works well with some oscillation. I attach here a video. https://drive.google.com/open?id=1sxK1SRzyOSfxDVE6_rPHDkVkjddDbnPS

R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 22:42, Monday 16 September 2019 (1642)Get code to link to this report
Sudden increase of BS OpLev high frequency spectrum

While working remotely on loops optimization, I noticed that the BS oplev high frequency spectrum suddenly increased. See attached picture. 

After some time (about ~20 min) it went back to the previous value. I suspect it might be due to an increase of environmental vibration maybe bought by the activation of some device.

Unfortunately I was not there to check. To be monitored.

Images attached to this report
1642_20190916154152_22.png
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 22:34, Monday 16 September 2019 (1641)Get code to link to this report
AC loops optimization

I spent some time to optimize the new damp loops, whitout DC control, to be used when we want to engage AA or dithering control.

In the attached file the spectra with open and close loops for BS, INPUT and END mirror are reported together with the poles/zeros of the filters.

In some cases, an optimization (espacially to reduce overshoot) is still necessary.

Non-image files attached to this report
R&D (FilterCavity)
Print this report.
YuhangZhao - 14:28, Sunday 15 September 2019 (1640)Get code to link to this report
Replacement of shelf beneath in-air bench

To separate devices we put beneath the in-air bench, Matteo prepared a new shelf, we replaced it. The situation is now like the attached figure.

Images attached to this report
1640_20190915072838_wechatimg515.jpeg
R&D (FilterCavity)
Print this report.
YuhangZhao - 14:24, Sunday 15 September 2019 (1639)Get code to link to this report
Test of galvo with NIM box

Yuhang and Yuefan

We tested the galvo for AA system.

We put a mirror after Faraday because there is still gate valve inside the vacuum tube. In this case(see attached figure 1), we could get a stable reflection. As mentioned by yuefan in the last entry, we had a problem of astigmatism in the beginning. We removed the second lens(-50mm lens), and then astigmatism disappeared.

Couldn't lock situation: If we found the light ERROR X or ERROR Y becomes light on(attached figure 2 and 3), it means we have a large offset in X or Y direction. So we couldn't lock loop and center beam.

Could lock situation: We found if error x is within 290mV and error y is within 730mV(attached figure 4 and 5), we could lock loop and center beam. But we found oscillation of around 30Hz or 1Hz after lock(attached figure 6 and 7).

The moment of lock is attached in figure 8.

Next step:

1. Get the real jittering beam and see if we could center beam.

2. Test galvo even with astigmatism.

Images attached to this report
1639_20190915072330_wechatimg514.jpeg 1639_20190915072341_wechatimg520.jpeg 1639_20190915072352_wechatimg519.jpeg 1639_20190915072408_wechatimg518.jpeg 1639_20190915072415_wechatimg517.jpeg 1639_20190915072425_wechatimg516.jpeg 1639_20190915072434_wechatimg521.jpeg 1639_20190915072442_wechatimg522.jpeg
R&D (FilterCavity)
Print this report.
YuefanGuo - 11:56, Sunday 15 September 2019 (1638)Get code to link to this report
Repeat the AA telescope simulation according to the result in entry 1598

Yuefan and Yuhang

Last Friday, for better understanding to beam situation, we decided to check again the simulation.

According to the beam size measured by Yuhang and Aritomi-san a while ago after the 500mm lens, we could get the original beam waist size and position of the reflection green beam. The beam waist is 2.5m before the 500mm lens, and the beam waist size is 440um which is different from what we expected (~1mm) and mearsured very long time ago by Yuhang and Marc(1008um).

Anyway then we checked the telescope I already designed with this beam, the result didn't seem good. The quadrant position for gouy phase 0 and 90 degree didn't change a lot, but the size became 100um on NF quadrant and 2.5mm on FF, both are bad for our requirement.

To increase the beam size on the NF quadrant, we could change the 500mm to longer one, but then this will limited mainly by the space we have. We decided to check again the beam profile next week when we get the cavity back to decide if we need to change the design or not.

Another thing we have to check is the astigmatism of the beam, because during the installation we found that the second lens which has a focal length of 50mm could induce a lot of astigmatism on the FF quadrant even we could not see by eyes the beam is off center. And if the beam already has a very strong astigmatism as measured in entry 1598, the final beam on the FF quadrant will have a really ugly shape. So after we check the beam again next, if it is still the same, we would like to check the injection path to see where the astigmatism come from.

R&D (FilterCavity)
Print this report.
NaokiAritomi - 16:03, Friday 13 September 2019 (1636)Get code to link to this report
Loss inside PR chamber

[Aritomi, Yuhang, Yuefan, Eleonora, Matteo, Takahashi]

We opened PR chamber and measured BAB power in PR chamber. BAB power was fluctuating a bit and BAB power after dichoic mirror seems not reasonable, but anyway loss of dichroic mirror should be small. Loss of faraday in PR chamber is around 5% and other loss is less than 1%.

position BAB power (uW)
before viewport 127
after viewport 126.7
after faraday  120.3
after dichroic mirror 121

Then we put mirror after dichroic mirror and reflected BAB into faraday and maximized the reflection. BAB injection inside PR chamber is 131uW and BAB reflection is 116uW. The round trip loss is 11.5%. 

KAGRA MIR (Polarization)
Print this report.
SimonZeidler - 13:51, Friday 13 September 2019 (1635)Get code to link to this report
Reconfiguration of the PCI to take polarization maps

Matteo, Simon

Yesterday, we finished to restart the translation stage after the power outage the day before.

After making sure that everything is working properly, we started to reconfigure the PCI bench for the polarization measurements:

  • placing ND-filters in the beam-path before the shutter (-> order ~ 2)
  • placing a Vantablack beam-dump at the same area to cover the ND-filter reflection
  • putting a PBS after the first lens of the telescope
  • putting a QWP  and a HWP after the PBS
  • placing a razor beam-dump before the second lens of the telescope to cover a ghost-beam coming from the ND-filters
  • putting a 1" lens (f=150mm) together with a beam-tube on the detector stage
  • putting a second PBS so that the beam-focus is ~ 2 cm behind the PBS
  • placing a beam-dumb behind the blind-side of the PBS to cover back-scattering
  • putting each a PSD (position sensitive detector) 2 cm behind the S-pol and P-pol side of the PBS
  • Connecting the PSDs to the Lock-Ins and fine-tune the position of the PSDs with an oscilloscope (use the sum-output for the Lock-Ins and the X/Y output for the fine-tuning; beam should be centered)

After the setup is done, we calibrated the QWP and HWP in an iterative manner to produce a maximum in P-pol.
The data on the calibration are as follows:
S-pol PSD maximum voltage: 285 mV
S-pol PSD minimum voltage: 0.77 mV

P-pol PSD maximum voltage: 349 mV
P-pol PSD minimum voltage: 0.2 mV

That is comparable to what we got before.

(Note, that these number might be imrovable as we didn't spent so much time in adjusting the PBS position.)

We started a map of the Shinkosha TAMA#2 sample during the night. The results can be seen from the attached figures.
The polarization is relatively homogeneous. However, there are two main areas with different polarization-angles on two sides which leaves the center of the sample with the largest gradient in the polarization angle.
It is noteworthy that the distribution of polarization angles remembers a little bit on the 3-fold splitting visible in the absorption maps (see the elog entry 1633).

Images attached to this report
1635_20190913065128_map20190912pol.png 1635_20190913065132_dist20190912pol.png
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 13:47, Friday 13 September 2019 (1634)Get code to link to this report
PR chamber closed. Vacuum restarted.

[Takahashi, Yuefan, Yuhang, Eleonora]

This morning with the help of Takahashi-san we closed PR chamber and restarted the vacuum. We will wait the vacuum to reach 10-6 torr (at BS vacuum monitor) to open the main gatevalve between BS and INPUT.

I upload some pictures taken to PR chamber. 

Images attached to this report
1634_20190913064437_pr1.jpeg 1634_20190913064442_pr2.jpeg 1634_20190913064447_pr3.jpeg 1634_20190913064452_pr4.jpeg 1634_20190913064500_pr5.jpeg 1634_20190913064505_pr6.jpeg 1634_20190913064512_pr7.jpeg 1634_20190913064517_pr8.jpeg 1634_20190913064529_pr9.jpeg 1634_20190913064546_pr10.jpeg
KAGRA MIR (Absorption)
Print this report.
SimonZeidler - 13:23, Friday 13 September 2019 (1633)Get code to link to this report
Results of TAMA-sized Sapphire Sample#2 Absorption Maps

Matteo, Simon

Attached to this report are the maps and distribution histograms of the absorption coefficient from the Shinkosha TAMA-sized TM #2.
As can be seen, we have now 7 different XY maps throughout the entire bulk. The YZ-map has a larger expansion in Y.

Although inhomogeneously distributed, the absorption is relatively low (~ 30 ppm/cm in the mean with some peaks exceeding 50 ppm/cm).
Please note that the colorscale is the same for all maps!

Images attached to this report
1633_20190913062217_map20190910xy46.png 1633_20190913062223_map20190909xy51.png 1633_20190913062227_map20190909xy56.png 1633_20190913062231_map20190906xy61.png 1633_20190913062237_map20190907xy66.png 1633_20190913062241_map20190907xy71.png 1633_20190913062245_map20190907xy76.png 1633_20190913062251_dist20190910xy46.png 1633_20190913062254_dist20190909xy51.png 1633_20190913062258_dist20190909xy56.png 1633_20190913062302_dist20190906xy61.png 1633_20190913062309_dist20190907xy66.png 1633_20190913062313_dist20190907xy71.png 1633_20190913062316_dist20190907xy76.png
R&D (FilterCavity)
Print this report.
YuefanGuo - 01:40, Friday 13 September 2019 (1632)Get code to link to this report
AA telescope installed according to the simulation

Yuefan and Yuhang

This morning we installed everything for the AA telescope before opening the chamber.

The final beam height is 225mm from the bench. All the optics position are only followed the design, we didn't have time to connect all the cables to check any signal. The second lens is installed on the rail, galvo and quadrants are all installed on the height changable posts.

The FF path is well aligned, the beam is hitting in the center of the quadrant. But the NF path, we still need to adjust a bit the mirrors of the galvo, to make the output beam on the same height and without any angle. It was not very easy to do because the position of the galvo is almost in the center of the bench.

The detail pictures are in the attachment. I found it was difficult to take a picture that shows all the components....I should also have taken some pictures from the other side of the bench, I will put them later...

Images attached to this report
1632_20190912183900_aatele1.jpg 1632_20190912183910_aatele2.jpg
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 19:58, Thursday 12 September 2019 (1631)Get code to link to this report
PR and BS chamber opened

[Takahashi, Aritomi, Yuefan, Matteo,Yuhang, Eleonora]

Today with the help of Takahashi-san we opened the BS and PR chambers.

First goal was to check a possible touching of the intermediate mass.

PR chamber

We found the intermediate mass was almost touching the magnet cage. Takahashi-san adjusted the cage position to leave more space between the cage and the intermediate mass. From the spectra taken before and after there is no evidence of improvement (pic1), but the spectra after the change have been taken with the chamber still open. So we will repet the measument in vacum to confirm the results.

BS chamber 

BS intermediate mass seems fine.

The second activity was the assesment of losses inside PR chamber, from viewport, Faraday, dichroic mirror etc. Aritomi and Yuhang will report about that.

We closed the BS chamber and left the PR one open with the cleanboth. We wll close it tomorrow moring.

Images attached to this report
1631_20190912125830_bsprintermediatemass.png
General (General)
Print this report.
EleonoraCapocasa - 10:41, Thursday 12 September 2019 (1630)Get code to link to this report
TAMA blackout due to a lightning storm

Yestarday afternoon at ~17.15 we had a blackout in TAMA due to a severe lightning storm. The power came back after ~10 minutes. 

We spent the rest of the afternoon recovering the system. For the filter cavity everything seems fine. Takahashi-san came to help with vacuum system recovery.

R&D (Cryogenic)
Print this report.
SatoshiTanioka - 20:16, Wednesday 11 September 2019 (1629)Get code to link to this report
Comment to PDH lock trial (Click here to view original report: 1621)

I re-tried lock, but could not.
At first, I adjusted the offset of servo to set the monitored error signal to zero.
Also I minimize the gain of servo.
I tuned the laser frequency around the resonance, and turned on the integrator.

I tried several times with different conditions (inverted or not, and so on..), however, could not lock...
There may be some problems such as high finess, large gain, etc...

Actually, the efficiency of laser PZT is 50MHz/V, and the slope of error signal is 4.2e-7V/Hz.
Therefore, it needs to optimize the gain of the servo.

R&D (FilterCavity)
Print this report.
NaokiAritomi - 18:56, Wednesday 11 September 2019 (1628)Get code to link to this report
Recovery of green alignment into OPO

I found that green into OPO is completely misaligned. I recovered the alignment by measuring parametric amplification. Here is the information of parametric amplification of BAB.

green power (mW) OPO temperature (kOhm) p pol PLL (MHz) Vmax (V) Vmin (V) parametric gain
0 7.2 305 0.112 0.112 1
40 7.185 165 1.82 0.04 16.3
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 16:16, Wednesday 11 September 2019 (1627)Get code to link to this report
Installation of optics for another AOM path

I installed the lens (f=50mm), QWP, and mirror for another double-pass AOM.
They are roughly located to compose a cat's eye retroreflector, but not tuned yet.

R&D (Cryogenic)
Print this report.
SatoshiTanioka - 16:10, Wednesday 11 September 2019 (1621)Get code to link to this report
PDH lock trial

I tried PDH lock with a servo borrowed from TAMA, but could not lock.
At that time I found the laser power was larger or smaller than usual.
I monitored output voltage from the servo, and it was about 14V which saturated PZT frequency tunig range.
I played with gain, but it did not affect the output.

Actually, I did not adjust the offset before turnig on the integrators.
So I will re-try the lock as Matteo suggested; tune the offset first, then turn on the integrators with smaller gain around the resonance.

Comments related to this report
SatoshiTanioka - 20:16, Wednesday 11 September 2019 (1629)

I re-tried lock, but could not.
At first, I adjusted the offset of servo to set the monitored error signal to zero.
Also I minimize the gain of servo.
I tuned the laser frequency around the resonance, and turned on the integrator.

I tried several times with different conditions (inverted or not, and so on..), however, could not lock...
There may be some problems such as high finess, large gain, etc...

Actually, the efficiency of laser PZT is 50MHz/V, and the slope of error signal is 4.2e-7V/Hz.
Therefore, it needs to optimize the gain of the servo.

R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 13:46, Wednesday 11 September 2019 (1626)Get code to link to this report
Real model update for AA -- part 2

Simulink model and MEDM screen for AA is now completed and ready to be tested with quadrants.

Note: while restarting the model last sunday I was not able to restore the screeshot file taken just before the restart. I didn't find any problem with the command but some of the epics channels remained blank. I check the file and see that some of that channels ware actually missing. I don't know why it happend. Anyway I restored a previous snapshot file. I restarted the model and restored snapshot several times after that and this problem never showed up again.

Images attached to this report
1626_20190911063801_aamodel.png 1626_20190911063807_aamodel2.png